Sign In | Join Free | My entremaqueros.com
China Xi'an Xinyi Instrument Technology Co., Ltd. logo
Xi'an Xinyi Instrument Technology Co., Ltd.
Verified Supplier

1 Years

Home > Pressure Calibrator Pump >

60Bar Hand Held Pneumatic Calibration Pump Clean Air Pressure Pump

Xi'an Xinyi Instrument Technology Co., Ltd.
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now

60Bar Hand Held Pneumatic Calibration Pump Clean Air Pressure Pump

Brand Name : XINYIYB

Model Number : YFP-60

Place of Origin : China

MOQ : 1PC

Payment Terms : L/C,D/P,D/A,T/T,Western Union

Supply Ability : 100pcs/month

Delivery Time : 5-8 work days

Packaging Details : Carton

Product name : Hand Held Pneumatic Pump

Type : Pneumatic Calibration Pump

Pressure range : -0.095 ~6.0MPa

Medium : clean air

Operating temperature : 0- 50℃

Contact Now

60 Bar Hand-Held Pneumatic Calibration Pump Clean Air Pressure Pump

Hand pump pressure transmitter calibrator is a portable pressure pump. This pressure transmitter calibrator is for field installation of instrument calibration in the industrial process and control of indoor use all kinds of pressure gauges.This hand pump pressure calibrator also provide pressure source correction table for pressure measuring instrument manufacturer. The handheld pressure calibrator for all kinds of pressure transmitter within every 60 Mpa,Pressure sensor and all kinds of pressure gauges Check all available hand pressure pump calibrator

Specification

item

value

Type

Pressure pump

Warranty

1 Year

Customized support

OEM, OBM, ODM

Brand Name

XINYI

Model Number

YFP-60

Product name

Hand held pneumatic pump

Pressure range

-0.095 ~6.0MPa

Medium

clean air

Type

pneumatic

Operation method

Pressure verification First, connect the tested meter to the output of the tested meter, and connect the standard meter to the output of the standard meter. Rotate the switch valve knob to push it out, and turn the pressure adjustment knob counterclockwise to close the pressure relief valve. Use the pressure handle to apply pressure to around 0.5-4MPa, and use the pressure adjustment knob to apply pressure to the desired level. When releasing pressure, simply open the pressure relief valve.

Vacuum calibration First, connect the tested meter to the output of the tested meter, and connect the standard meter to the output of the standard meter. Rotate the switch valve knob and press it in, then turn the pressure adjustment knob clockwise to close the pressure relief valve. Use the pressure handle to draw pressure to the desired vacuum value. When releasing pressure, simply open it.

Note

1. The device should be used within the rated pressure range as much as possible and should not be used beyond the range.

2. The pressure/vacuum conversion must be carried out in a non pressure state.

3. The calibration table should be placed on a stable operating platform to avoid entering environments with corrosive gases and dust particles.

4. If not used for a long time, the equipment should be stored in a dry, non corrosive gas environment;

5. Leverage pressure, apply even force, do not use force when closing or opening the valve.

6. Regularly clean the impurities inside the filter to prevent them from affecting the service life of the equipment.

60Bar Hand Held Pneumatic Calibration Pump Clean Air Pressure Pump

60Bar Hand Held Pneumatic Calibration Pump Clean Air Pressure Pump

60Bar Hand Held Pneumatic Calibration Pump Clean Air Pressure Pump


Product Tags:

60Bar Pneumatic Calibration Pump

      

Hand Held Pneumatic Calibration Pump

      

Clean Air Pressure Pump

      
 60Bar Hand Held Pneumatic Calibration Pump Clean Air Pressure Pump Manufactures

60Bar Hand Held Pneumatic Calibration Pump Clean Air Pressure Pump Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: Xi'an Xinyi Instrument Technology Co., Ltd.
*Subject:
*Message:
Characters Remaining: (0/3000)